Recall Z-1424-2021 ยท JSO Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar

Remel Inc

Remel Haemophilus Test Medium (150 mm) 10/PK, REF 04033 Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative procedures by the Clinical and Laboratory Standards Institute (CLSI) as the growth medium for antimicrobial disk diffusion tests with Haemophilus species.

DateMar 10, 2021
StatusOpen, Classified
ReasonWhen tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure.
Root causeUnder Investigation by firm
ActionThe firm, ThermoFisher Scientific, sent an "Medical Device Field Action Recall" letter dated 3/10/2021 to its consignees on 03/10/2021. The letter described the product, problem and actions to be taken. The consignees were instructed to do the following: Review the reported test results should be determined by the appropriate technical expert. Please keep this notification on file. This notice needs to be passed on to all who need to be aware within your organization or to any organization where the potentially affected products have been transferred. If you have any questions, please cont
Quantity408 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K892928HAEMOPHILUS TEST MEDIUM (AGAR)
Remel Co., Jun 13, 1989