Recall Z-1282-2021 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Hitachi Medical Systems America Inc

Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained personnel (doctor, Sonographer, etc.) for the diagnostic ultrasound evaluation of Fetal, Abdominal, Intra-operative (Spec.), Intraoperative (Neuro.), Laparoscopic, Pediatric, Small Organ (Spec.), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-esoph. (non-Card.), Musculo-skel. (Convent.), Mu

DateMar 1, 2021
StatusTerminated
ReasonArietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Root causeSoftware design
ActionOn 3/1/2021, Hitachi issued an Urgent Medical Device Correction notice to customers for the Arietta 850 system operating with software version 4.0.2 that utilizes SWE function with the C252 probe. The R&D department of Hitachi Ltd. (Manufacturer) has identified an error in the focus point and transducer aperture settings in the SWE function at the manufacturer facility. This error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Quantity1
DistributionUS Nationwide distribution in the state of TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183456Aloka Arietta 850
Hitachi Healthcare Americas, Jan 28, 2019