Recall Z-1313-2021 ยท DWA Control, Pump Speed, Cardiopulmonary Bypass

Medtronic Perfusion Systems

Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1, R560BCS1. Used in cardiopulmonary bypass (CPB) procedures.

DateMar 3, 2021
StatusOpen, Classified
ReasonAn electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use.
Root causeComponent change control
ActionIn the United States (US), beginning 03-Mar-2021, a written Urgent Medical Device Correction letter was mailed via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete and return a confirmation certificate, confirming they have received the notification.
Quantity387 devices (93 US)
DistributionWorldwide distribution. US Nationwide including Puerto Rico, Brazil, Brunei Darussalam, Canada, China, Colombia, Croatia, Ecuador, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Korea, Mexico, Morocco, New Zealand, Russian Federation, Spain, Taiwan, Thailand, United Kingdom, and Viet Nam