ClearanceBrief
Recall Z-1375-2021 ยท FRN Pump, Infusion

Pacific Medical Group Inc.

Aftermarket Alaris Infusion Pump Module Model 8100 Front Bezel, installed by Pacific Medical Group (DBA Avante Health Solutions)

DateMar 24, 2021
StatusOpen, Classified
ReasonAftermarket front bezel components were installed during service/repair, not by the original manufacturer, within infusion pump modules. Over time, the posts on the front bezel component may crack or separate. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
Root causeComponent design/selection
ActionOn 03/24/2021, Medical Device Recall Notifications were emailed to customers. Customers were asked to take the following actions: If you have affected product, remove these devices from clinical use and immediately quarantine them until the impacted parts can be returned to the recalling firm; contact Customer Service 800-449-5328, pm.recallapm@avantehs.com. The recall part number is RCL2021-823-01. Once returned, the affected device will be evaluated and inspected and where the component under recall is identified, it will be replaced by our organization with the OEM component and then retur
Quantity2452
DistributionDistribution US nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133532ALARIS SYSTEM WITH GUARDRAILS SUITE MX
Carefusion 303, Inc., Aug 21, 2014