Recall Z-1288-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Interventional Fluoroscopic X-Ray System - Product Usage: developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

DateFeb 26, 2021
StatusTerminated
ReasonSiemens has become aware of a potential issue in which the imaging system UPS sporadically forces a shutdown of the imaging system PC without a true power supply problem.
Root causeComponent design/selection
ActionOn 2/26/2021, Siemens issued an Urgent Medical Device Correction notice to in inform customers of a possible hazard to patients, operators, or other persons and equipment due to a hardware error which affects ARTIS icono /pheno systems in combination with imaging system UPS (Uninterrupted Power Supply) option. This potential hardware issue has no influence on the previous treatment/diagnosis of patients. However, in case of error, planned procedures may have to be terminated and performed on an alternative x-ray system. Based on our investigation, you can continue to safely use your system.
Quantity29 systems
DistributionUS Nationwide distribution in the states of MI, KY, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193326ARTIS Icono
Siemens Medical Solutions USA, Inc., Apr 16, 2020