Recall Z-1273-2021 ยท JAK System, X-Ray, Tomography, Computed

Philips Healthcare (Suzhou) Co., Ltd.

Philips CT systems are advanced continuous-rotation computed tomography systems suitable for a wide range of computed tomographic (CT) applications - Product Usage: intended for use in the head and whole body.

DateFeb 24, 2021
StatusTerminated
ReasonPhilips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.
Root causeProcess control
ActionOn 2/18/2021, Philips issued an Urgent Medical Device Correction because the firm has detected an issue with some actuators used in the couch of Incisive CT and Access CT systems, which could result in the couch unexpectedly moving slowly downward to its limit position.
Quantity109 systems
DistributionWorldwide distribution - US Nationwide distribution in the states of KY, FL, St. Croix, and the countries of Australia, Bolivia, China, Germany, Hungary, Italy, Japan, Republic of Korea, Latvia, Palestine, Puerto Rico, Romania, Russia, Saudi Arabia, Senegal, Serbia, Spain, Sweden, Thailand, Uzbekist
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180015Philips Incisive CT
Philips Healthcare (Suzhou) Co., Ltd., Mar 20, 2018