ClearanceBrief
Recall Z-1175-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Canon Medical System, USA, INC.

System INFX-8000C

DateFeb 5, 2021
StatusOpen, Classified
ReasonThe x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.
Root causeSoftware design
ActionA customer notification letter was sent to affected customers, which includes instructions to (1) manually operate the diaphragm blades of the collimator to adjust the X-ray irradiation field to the appropriate irradiation field position, (2) when adjusting the diaphragm blades, you can find this problem by checking the position of the diaphragm blades with respect to the image receiving surface displayed on the screen, (3) contact their Canon Service representative if they find any abnormalities regarding this issue in the operation of the Infinix System and if they have any questions. And in
Quantity272 in total
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152697INFX-8000C, V6.20
Toshibamedical Systems Corporation, Jan 15, 2016