Recall Z-1284-2021 ยท MXA System, Test, Breath Nitric Oxide

CAIRE DIAGNOSTICS INC

CAIRE DIAGNOSTICS FenomPro REF 900-0001 SN ***** - Product Usage: a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. Fenom Pro should not be used in critical care, emergency care or in anesthesiology.

DateFeb 26, 2021
StatusTerminated
ReasonErroneously high FeNO levels which may contribute to premature discontinuation of effective asthma treatment and persistent underlying disease. Two issues affecting the FeNO readings: 1) Drift in the calibration gases and 2) Software error. Each of these issues can cause FeNO scores/results to be erroneously high.
Root causeProcess control
ActionBetween 02/21-26/2021, the firm contacted its customers via email/phone and Zoom to inform them that two calibration errors were noted for the Asthma Monitor regarding, 1) drift in calibration gases and 2) a software error that may cause a high reading indicating that a therapeutic action should be taken. This may result in prescription of unneeded medication to the patient. On 02/26/2021, the firm sent a follow-up "Urgent - Medical Device Recall" Notification further instructing the customers to: 1. Discontinue use of the device. Make all personnel who use it aware of this recall. 2. Contact
Quantity381 units (171 units U.S. and 210 units O.U.S.)
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of CA, CO, FL, TX, and WI. In the countries of Belgium, Canada, Croatia/Hrvatska, France, Germany, Greece, Hong Kong, India, Philippines, Spain, Spain and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182874Fenom Pro Nitric Oxide Test
Spirosure, Inc., Feb 13, 2019