ClearanceBrief
Recall Z-1302-2021 ยท ILG Stocking, Elastic

Cardinal Health 200, LLC

In-View Clear Leggings Bulk Non-Sterile

DateFeb 25, 2021
StatusTerminated
ReasonCardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.
Root causeNonconforming Material/Component
ActionA Recall notification letter titled, "URGENT MEDICAL DEVICE RECALL" was sent to consignees on 02/25/2021 via FedEx priority overnight. The consignee is requested to follow the following instructions: 1. CHECK all storage and usage locations to confirm whether they have any units of the affected product codes and lot numbers listed in Attachment 1. 2. SEGREGATE and QUARANTINE all impacted on-hand product. 3) RETURN the enclosed acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity
Quantity21343 units
DistributionDistributed nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Distributed also to Australia and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.