Recall Z-1333-2021 ยท KPA Slide Stainer, Automated

Biocare Medical, LLC

BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containing software version 3.5.3.1

DateFeb 25, 2021
StatusTerminated
ReasonTheir is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations.
Root causeSoftware Design Change
ActionOn 2/25/2021, BIOCARE Medical emailed and "Medical Device Recall" Notification to all affected consignees. In addition to informing consignees about the recall, the firm requested consignees take the following actions: 1. Do not use the product in the Multi Batching mode. 2. Do not interrupt a run during a Single Batch mode run or access the working deck while the instrument is in operation. 3. Always run positive controls per CAP and CLIA guidance to avoid the potential for false negative result from contaminated reagents. 4. If the malfunction does occur: " Do not open the cover and acces
Quantity72 units
DistributionUS: AL, AZ, CA, CT, FL, IL, MA, MD, MN, NC, OH, OR, PA, WA, WY OUS: Germany, Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.