Recall Z-1254-2021 ยท MSX System, Network And Communication, Physiological Monitors

Mindray DS USA, Inc. dba Mindray North America

Mindray BeneVision Distributed Monitoring System( DMS) 1. p/n 115-051209-00 2. p/n 115-034414-00 and 115-034413-00 when used with the following software versions: 03.04.00 04.01.00 04.03.00 04.04.00 04.04.01

DateFeb 22, 2021
StatusOpen, Classified
ReasonBeneVision DMS may intermittently freeze and require a manual reboot after which normal operation resumes. If a freeze occur, patients monitored on a telemetry transmitter will no longer communicate data to the BeneVision DMS
Root causeSoftware design
ActionMindray issued URGENT Medical Device Correction letter on February 22, 2021 via certified mail, return receipt, signature required. Letter stated reason for recall, health risk and action to take: Should the BeneVision DMS freeze, perform a reboot of the system and normal monitoring will resume. If needed, Mindrays Technical Support team can assist and may be reached at (877) 913-9663 (Option 1), Monday through Friday, 8:30am - 5:30pm ET. A configuration change will be deployed by Mindrays Service team and is anticipated to take approximately 10-15 minutes per system. A representative from Mi
Quantity1938 systems
DistributionNationwide Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193391BeneVision Central Monitoring System
Shenzhen Mindray Bio-Medical Electronics Co., Ltd., May 21, 2020