Recall Z-1186-2021 ยท MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

Howmedica Osteonics Corp.

Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer.

DateFeb 10, 2021
StatusTerminated
ReasonStryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness mark (taper) at the bottom of the internal profile of the baseplate. The taper prevents the insert from locking on the baseplate.
Root causeUnder Investigation by firm
ActionOn 2/11/2021, Stryker issued Urgent Medical Device Recall notices to customers via UPS 2-day air stating Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized.
Quantity11 devices
DistributionDomestic: AZ, GA, MA, MI, MN, NH, NJ, OR, PA, SC, UT, WA, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123486TRIATHLON TRITANIUM TIBIAL BASEPLATE
Stryker, May 3, 2013
K141056TRIATHLON KNEE SYSTEM POSTERIOR STABILIZED COMPONENTS
Stryker Orthopaedics, Aug 5, 2014