ClearanceBrief
Recall Z-1269-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Canon Medical System, USA, INC.

Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000V Catheterization Table CAT-880B Canon CATHETERIZATION TABLE, MODEL CAT-880B INPUT 220V - Product Usage: indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.

DateDec 8, 2020
StatusOpen, Classified
ReasonSystem table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
Root causeComponent design/selection
ActionOn 01/12/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter via USPS return receipt mail to customers informing them that the catheterization tables may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose. Additionally, the manual operation of the tabletop longitudinal or lateral movement will become heavy or hard to move due to the loosening of the bolts. Customers are instructed/informed: a. Should the table top tilt during a procedure before the bolts are replaced and secured, to ensure the safety of the p
QuantityTotal of all Systems = 145 systems (U.S.)
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170909Infinix INFX-8000V, V6.40
Toshiba Medical Systems Coroporation, Jun 16, 2017