Recall Z-1200-2021 ยท MIJ Needle, Tumor Localization

Somatex Medical Technologies GMBH

Thread-like wire marker

DateFeb 3, 2021
StatusTerminated
ReasonThis recall has been initiated due to the risk of wire splintering after contact with a cautery device during the removal of a lesion. There is a potential risk that wire fragments are not removed during the procedure and stays inside the patient's body.
Root causeDevice Design
ActionOn 1/22/2021, Somatex issued an Urgent Medical Device Removal notice to customers via e-mail for Tuflex and Tuflex Premium marking wires. This recall has been initiated due to the risk of wire splintering after contact with a cautery device during the removal of a lesion.
Quantity196 pieces
DistributionDomestic: Mo, NJ, OH
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.