Recall Z-1289-2021 ยท KNW Instrument, Biopsy

Bard Peripheral Vascular Inc

BARD MARQUEE Disposable Core Biopsy Instruments XX.Xcm, XX.Xcm, XXg x XXcm REF XXXXXXX Note - for definition of the "X's refer to Tyvek Lid Label.pdf" uploaded in Associated Documents

DateJan 14, 2021
StatusTerminated
ReasonDue to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.
Root causeProcess control
ActionOn 02/11/2021, the firm sent an "UGENT MEDICAL DEVICE RECALL" Notification Letter via FedEx or email to customers/distributors informing them that the Recalling Firm has received reports that the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use and may results in (1) prolongation of the biopsy procedures as the user obtains a replacement device, (2) undershooting the lesion unsuspectedly, compromising the sample quality and potentially leading to a misdiagnosis (3) or if the penetration is o
QuantityTotal 131,876 kits (102,210 kits U.S. and 29,666 kits O.U.S.)
DistributionU.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY O.U.S.: Australia, Belgium, Canada, Europe, India, and Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133948BARD(R) MONOPTY(R) DISPOSABLE CORE BIOPSY INSTRUMENT, BARD(R) MAX-CORE(R) DISPOSABLE CORE INSTRUMENT
Bard Peripheral Vascular, Inc., Feb 21, 2014