Recall Z-1202-2021 ยท DOP Mass Spectrometer, Clinical Use

Thermo Fisher Scientific

Cascadion SM Clinical Analyzer, REF 99990000 SN Type: 9910

DateFeb 1, 2021
StatusTerminated
ReasonDue to software defect, under certain assay parameters, false Vitamin D results may be reported. The system reports a false result by not quantitating the correct analyte peak in sample chromatogram. The resulting sample would show a Vitamin D-concentration that is an unusual situation for human serum and plasma samples and would indicate a severe Vitamin D deficiency.
Root causeSoftware design
ActionOn 02/01/2021, the firm sent an "URGENT - MEDICAL DEVICE RECALL" Notification Letter to inform customers that the Recalling Firm has identified that there is a software anomaly in the Analyzer Software subcomponent called Suhaili Service (version 2.0.3 and earlier) that is affecting the Analyzer. Suhaili Service controls raw data processing based on assay configuration data from Sample Prep software. Processed results are sent to Sample Prep software for patient result reporting. This anomaly, as a result of this interface, has a potential risk of reporting wrong patient results for Vitamin D.
Quantity19 systems (1 system in U.S. and 10 system O.U.S.)
DistributionU.S.: Texas O.U.S.: Belgium, France, Germany, Italy, Spain, Switzerland and United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.