Recall Z-1263-2021 ยท DRC Trocar
Merit Medical Systems, Inc.
Heartspan Transseptal Needles, Catalog No. FND-019-02, Lot No. E1913645 consisting of an outer needle cannula and an inner stylet. The needle is comprised of flexible thin walled tubing with an ergonomic hub and stopcock attached to the proximal end. The stylet consists of a solid wire that when inserted in the needle protrudes beyond the distal tip of the cannula.
| Date | Jan 19, 2021 |
| Status | Terminated |
| Reason | The labeled needle tip curvature of specific lots of transseptal needles may not match the actual needle tip curvature. In the event that the incorrect needle is unknowingly used, an unintended anatomy puncture may occur, which is likely to require medical intervention. |
| Root cause | Process control |
| Action | Urgent Product Recall Notices were sent to consignees via FedEx overnight delivery beginning on January 19, 2021. The notification included product identification information, instructions to determine if any of the devices listed in the Customer Response Form (CRF) were within the consignees' faciltiies, quarantine them, and to discontinue distribution, and return product to Merit. Consignees were also requested to ensure that applicable personnel within their organizations were made aware of this field action; if the product has been further distributed to other facilities, institutions, or |
| Quantity | 448 Units |
| Distribution | U.S. Consignees: VA, NE, Washington DC, MI, NY, KY, CA, TX, MD, AL, GA, UT, LA, CA, CT, PA, WA, MT and NJ. No government consignees. Also distributed OUS. |
| FDA record | Open on fda.gov |