Recall Z-1181-2021 ยท BSZ Gas-Machine, Anesthesia

Getinge Group Logistics America, LLC

Anesthesia gas-machine - Product Usage: intended for use in administering anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with limited ability to breathe.

DateJan 29, 2021
StatusTerminated
ReasonMaquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine.
Root causeDevice Design
ActionOn 2/9/2021 Maquet Critical Care AB / Getinge issued an Urgent Medical Device Correction letter via FedEx for the Flow i anesthesia machine due to the potential for high FiCO2 values and/or a System Check Out failure due to an incomplete docking of the CO2 absorber to the patient cassette. The patients re-breathing without CO2 reduction could occur if the absorber is not sufficiently docked.
Quantity144 units
DistributionUS Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, MA, MI, MO, NJ, NY, OH, OR, PA, TX, VA, WA, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System
Maquet Critical Care AB, Mar 18, 2020