Recall Z-1153-2021 · GEI Electrosurgical, Cutting & Coagulation & Accessories

Bovie Medical Corporation

Velocity Disposable Bipolar Ablators - Product Usage: intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures.

DateFeb 1, 2021
StatusOpen, Classified
ReasonThere can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration.
Root causeDevice Design
ActionThe firm sent "URGENT MEDICAL DEVICE RECALL" (FA-2021-1) and response form to customers on February 1, 2021.
Quantity5894 units
DistributionUS Nationwide distribution in the states of CA and FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152777Bovie Disposable Bipolar Ablator
Bovie Medical Corporation, Jan 15, 2016
K161558Bovie 3.3mm Disposable Bipolar Ablator 90°, Aspirating and Non-Aspirating, Bovie 2.4mm Disposable Bipolar Ablator 55°, N