Recall Z-1173-2021 ยท DQK Computer, Diagnostic, Programmable
Siemens Medical Solutions USA, Inc
Sensis / Sensis Vibe systems with software version VD12 and equipped with HP Flex Pro-C PC. Diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies Model Numbers: Sensis Vibe Hemo 11007641 Sensis Vibe Combo 11007642 Sensis 10764561 Postprocessing Workstation 6648161 SIS Server 6648153
| Date | Jan 27, 2021 |
| Status | Terminated |
| Reason | System may sporadically freeze (lock-up) during operation or while being in an idle state, no longer possible to interact with the system |
| Root cause | Software design |
| Action | Siemens issued Urgent Medical Device Correction via Customer Safety Advisory Notice AX010/21/S dated 1/27/21distributed to all affected customers. Additionally, a solution to eliminate the root cause of this problem is being distributed via Update Instruction AX009/21/S to all affected customers. Letter states reason for recall, health risk and action to take: The freeze is a total lock-up of the system and requires a power cycle of the PC to recover. Please make sure that an alternative system can be used to continue treatment in such cases.Our service organization will contact you shortly t |
| Quantity | 37 US |
| Distribution | US Nationwide distribution in the states of AK, AR, CA, GU, IA, IL, MI, MO, NC, NE, OR, PA, PA, PR, SD, TN, TX, WA. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K150493 | Sensis Siemens Medi Cal Solutions, Inc., Jun 30, 2015 |