Recall Z-1173-2021 ยท DQK Computer, Diagnostic, Programmable

Siemens Medical Solutions USA, Inc

Sensis / Sensis Vibe systems with software version VD12 and equipped with HP Flex Pro-C PC. Diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies Model Numbers: Sensis Vibe Hemo 11007641 Sensis Vibe Combo 11007642 Sensis 10764561 Postprocessing Workstation 6648161 SIS Server 6648153

DateJan 27, 2021
StatusTerminated
ReasonSystem may sporadically freeze (lock-up) during operation or while being in an idle state, no longer possible to interact with the system
Root causeSoftware design
ActionSiemens issued Urgent Medical Device Correction via Customer Safety Advisory Notice AX010/21/S dated 1/27/21distributed to all affected customers. Additionally, a solution to eliminate the root cause of this problem is being distributed via Update Instruction AX009/21/S to all affected customers. Letter states reason for recall, health risk and action to take: The freeze is a total lock-up of the system and requires a power cycle of the PC to recover. Please make sure that an alternative system can be used to continue treatment in such cases.Our service organization will contact you shortly t
Quantity37 US
DistributionUS Nationwide distribution in the states of AK, AR, CA, GU, IA, IL, MI, MO, NC, NE, OR, PA, PA, PR, SD, TN, TX, WA.