Recall Z-1125-2021 ยท QKO Reagent, Coronavirus Serological

Confirm Biosciences Inc

Healgen COVID-19 IgG/IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) REF GCC)V-402a - Product Usage: is a lateral flow immunoassay intended for the qualitative detection and differentiation of IgM and IgG antibodies to SARS-CoV- 2 in human venous whole blood, plasma from anticoagulated blood or serum.

DateJan 29, 2021
StatusTerminated
ReasonDue to two (2) issues: 1) The firm distributed COVID-19 Rapid Test Cassettes outside of the Emergency Use Authorization to non-certified Clinical Laboratories. 2) COVID-19 Rapid Test Cassettes were not adequately validated for its intended use (fingerstick tests for in-home use/testing).
Root causeOther
ActionOn 01/29/2021, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Notification its customers and distributors informing them that Recalling Firm may have distributed products to facilities that do not meeting the conditions outline in the Emergency Use Authorization (EUA) which stipulates that only laboratories certified under Clinical Laboratory Improvement Amendments of 1988 (CLIA) to perform moderate or high complexity tests are authorized to receive and utilize the products. Use of these tests by non-certified laboratory facilities may result in false-positive test results and may lead t
Quantity563,840 cassettes
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY. The countri
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.