Recall Z-1306-2021 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Implantable cardioverter defibrillators (ICD) with cardiac resynchronization therapy (CRT-D): 1) Viva family: Viva XT CRT-D, Viva XT Quad CRT-D, Viva S CRT-D, Viva Quad S CRT-D and Viva Quad C CRT-D models; 2) Claria family: Claria MRI CRT-D and Claria MRI Quad CRT-D models; 3) Amplia family: Amplia MRI CRT-D and Amplia MRI Quad CRT-D models; 4) Compia family: Compia MRI CRT-D model; 5) Brava fa

DateMar 27, 2021 (initiated Feb 3, 2021)
StatusOpen, Classified
ReasonA small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Root causeDevice Design
ActionThe firm, Medtronic, issued an "Urgent Medical Device Correction" letter dated February 2021 to its consignees in the United States via UPS 2-day delivery. The letter described the product, problem and actions to be taken. Medtronic recommends the customers do the following: Continue normal follow-up per local clinical protocol. o Recognize that patients who require significant pacing support and high voltage therapy have the lowest risk for this issue - See Appendix A for additional details. o Where possible, take advantage of the CareLink" home monitoring system and the wireless low batte
Quantity238393 units
DistributionWorldwide Distribution: US (nationwide) Including Puerto Rico and countries of: Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, C