Recall Z-1305-2021 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Implantable cardioverter defibrillators (ICD): 1) Evera family: Evera XT DR, Evera S DR, Evera MRI XT DR, Evera MRI S DR, Evera XT VR, Evera S VR and Evera MRI S VR models; 2) Visia family: Visia AF ICD, Visia AF S ICD, Visia AF MRI ICD, and Visia AF MRI S ICD models. Implantable Cardioverter Defibrillators (ICDs) are implantable medical devices that automatically detect and treat episodes of

DateMar 27, 2021 (initiated Feb 3, 2021)
StatusOpen, Classified
ReasonA small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Root causeDevice Design
ActionThe firm, Medtronic, issued an "Urgent Medical Device Correction" letter dated February 2021 to its consignees in the United States via UPS 2-day delivery. The letter described the product, problem and actions to be taken. Medtronic recommends the customers do the following: Continue normal follow-up per local clinical protocol. o Recognize that patients who require significant pacing support and high voltage therapy have the lowest risk for this issue - See Appendix A for additional details. o Where possible, take advantage of the CareLink" home monitoring system and the wireless low batte
Quantity290086 units
DistributionWorldwide Distribution: US (nationwide) Including Puerto Rico and countries of: Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, C