ClearanceBrief
Recall Z-1207-2021 ยท KWQ Appliance, Fixation, Spinal Intervertebral Body

RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)

CervAlign Anterior Cervical Plate System-Cervalign, 1 Level, 16mm Material Number: 66-116 - Product Usage: intended for anterior cervical fixation (C2-C7) for the following conditions: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e. fracture or dislocation), spinal

DateJan 15, 2021
StatusOpen, Classified
ReasonLocking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
Root causeUnder Investigation by firm
ActionPioneer Surgical Technology, Inc. (DBA RTI Surgical) initiated email notification on 15 January 2021 via the Distributor. End Users notified via Distributor by First Class Mail. Letter states reason for recall, health risk and action to take: Remove from all user sites and inventory locations the CervAlign Anterior Cervical Plates listed in the attached distribution list. Kindly return CervAlign Anterior Cervical Plates RTI Surgical along with the appropriate forms. RTI Surgical is working on corrective actions to prevent these events from reoccurring. We do not know at this time how long th
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.