Recall Z-1148-2021 ยท FCX Insufflator, Automatic Carbon-Dioxide For Endoscope

Northgate Technologies, Inc.

Steris CO2MPACT Endoscopic Insufflator system, Catalog Number 72-00310-0 / 710300 - Product Usage: designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope.

DateJan 15, 2021
StatusTerminated
ReasonAn internal tubing component in the flow path was made with incorrect material which was not tested for use with this device.
Root causeNonconforming Material/Component
ActionThe firm initiated the recall on 01/15/2021 by email to its direct consignees. The notice requested the consignee cease distribution or use and return the product to NTI. It also directed the consignee to notify their customers if the product was further distributed.
Quantity45 units
DistributionUS Nationwide distribution in the states of FL and OH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111648BRACCO DIAGNOSTICS, INC. CO2MPACT ENDODCOPIC INSUFFLATOR
Bracco Diagnostic, Inc., Nov 3, 2011