Recall Z-1149-2021 ยท JEA Table, Surgical With Orthopedic Accessories, Ac-Powered

Steris Corporation

OT1000 Series Orthopedic Surgical Tables

DateJan 21, 2021
StatusTerminated
ReasonOn January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
Root causeProcess control
ActionIn January, 2021, Steris will issue an Urgent Medical Device Correction notice to customers via letter for the OT1000 Series Orthopedic Surgical Table. Steris has identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
Quantity272 tables total (Domestic: 227 tables, Foreign 45 tables)