Recall Z-1136-2021 ยท DIP Enzyme Immunoassay, Diphenylhydantoin
Ortho-Clinical Diagnostics, Inc.
VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.
| Date | Jan 14, 2021 |
| Status | Terminated |
| Reason | A deformity affecting a specific lot of VITROS Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays. |
| Root cause | Under Investigation by firm |
| Action | On 01/18/2021, Ortho Clinical Diagnostics issued an Urgent Product Correction Notification via e-mail due to a a deformity in a specific lot of VITROS Chemistry Products VersaTips. The affected product is to be no longer distributed and discarded. |
| Quantity | 4,062 cartons of 1,000 tips (Totaling 4,062,000 tips) |
| Distribution | Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the country of Canada. |
| FDA record | Open on fda.gov |