ClearanceBrief
Recall Z-1431-2021 ยท KOD Catheter, Urological

C.R. Bard, Inc.

Bard Clean-Cath Intermittent Catheter, 12 French female length (catheter, urological) - Product Usage: for urological use only.

DateJan 11, 2021
StatusTerminated
ReasonThe device catheter tip was cut off (no tip) and the product packaging was not fully sealed.
Root causePackaging process control
ActionAll customers will be contacted via letters sent via Fed Ex. The acknowledgement form included with the letter that is mailed or faxed back to BD will be reconciled with the customer list. BD will contact 100% of our customers that have not responded to the communication to verify acknowledgement. BD is requesting that customers discard affected components.
Quantity9,800 Units
DistributionWorldwide distribution - US Nationwide distribution and the countries of Australia, Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K802665TOUCHLESS CLEAN-CATH
Diamed Division Ill. Tool Works, Inc., Nov 12, 1980