Recall Z-1107-2021 ยท LSW Anti-Dna Antibody, Antigen And Control

Phadia US Inc

EliA dsDNA Well - Product Usage: intended for the In-Vitro quantitative measurement of IgG antibodies directed to dsDNA in human serum and plasma as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE). Material Number: 14-5500-01

DateFeb 11, 2021 (initiated Dec 14, 2020)
StatusTerminated
ReasonDecreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).
Root causeUnder Investigation by firm
ActionPhadia US Inc. (Initial Importer/Distributor) issued notification to consignees via phone and FedEx overnight delivery Dec. 17-18 . Letter states reason for recall, health risk and action to take: Customers are asked to perform the following steps: 1.Scrap EliA dsDNA Well lot BFA3Y/0142 and order a replacement free of charge 2. Assess test results from EliA dsDNA Well lot BFA3Y/0142 and determine if retesting of samples is warranted in accordance with internal operating procedures. 3. Complete the Acknowledgement Form in the Medical Device Field Safety Notice and return the response to the c
Quantity365 kits US
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072393ELIA DSDNA IMMUNOASSAY, MODEL 14-5500-01, ELIA ANA CONTROL, MODEL 83-1004-01
Phadia US, Inc., Dec 7, 2007