Recall Z-1117-2021 ยท JXG Shunt, Central Nervous System And Components

Medtronic Neurosurgery

Medtronic PS Medical Cerebro Spinal Fluid (CSF)-Flow Control Valves for the following CFNs: 22011 L- SHUNT 22011 L INTL ONLY ULTRA SM LOW 22011 LL - SHUNT 22011 LL INTL ONLY ULTRA SM L-L 22011 M - SHUNT 22011 M INTL ONLY ULTRA SM MED 22017 B-L - SHUNT 22017 B-L ASSY US PER 90CM SM LOW 22017 B-M - SHUNT 22017 B-M ASSY US PER 90CM SM MED 22017B-LL - SHUNT 22017B-LL ASSY US PER 90CM SM LL 23093 -

DateJan 11, 2021
StatusOpen, Classified
ReasonPotential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus
Root causeProcess control
ActionOn January 11, 2011, the firm issued an "URGENT: Medical Device Correction" notification to all affected consignees. In addition to informing consignees about the correction, the firm ask consignees to take the following actions; 1. Be aware that radiopaque marker visualization with impacted devices under radiographic imaging may be insufficient to determine the pressure level and/or flow direction. In these situations, alternative methods for this determination should be considered, if deemed necessary per clinical discretion. 2. Share this information within your organization to healthcare
Quantity20,817 units
DistributionWorldwide distribution. US nationwide, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Plurinational State Of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Cos
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K792007CSF-VENTRICULAR CATHETER, STANDARD DES
Pudenz-Schulte Medical Research Corp., Jan 16, 1980
K831678P-S MEDICAL CSF-FLOW CONTROL VALVE-BUR
Pudenz-Schulte Medical Research Corp., Jun 24, 1983
K841442P-S MEDICAL CSF-FLOW CONTROL VALVE-
P-S Medical, Apr 19, 1984
K873247CSF-FLOW CONTROL VALVE, ULTRA SMALL
Pudenz-Schulte Medical Research Corp., Sep 18, 1987
K951258PS MEDICAL BIOGLIDE CSF-FLOW CONTROL SHUNT KIT
Pudenz-Schulte Medical Research Corp., Sep 18, 1995