Recall Z-0940-2021 ยท QJR Reagents, 2019-Novel Coronavirus Nucleic Acid

Biomeme, Inc.

Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human clinical specimens. The RT-PCR test runs on a Franklin Real-Time PCR Thermocycler Device (Reference/Catalog Number 1000003,1000018) with the Software Controller version 6.4.3. The cycler has a "baseboard" unit with firmware and associated software

DateJan 5, 2021
StatusTerminated
ReasonUsers cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.
Root causeSoftware design
ActionOn 01/05/2021 Biomeme, Inc., issued an Urgent Device Recall Correction for the latest build of the Franklin Real-Time PCR Thermocycler, which affects some devices built since September 21st, 2020. This software correction is to fix an issue where the device is unable to get past the beginning stages of the test.
Quantity147 systems
DistributionDomestic: FL, TX, IN, KY, OH, NY, CA, PA, MD. Foreign: Canada, Zimbabwe.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.