Recall Z-0875-2021 ยท QKO Reagent, Coronavirus Serological

Beckman Coulter Inc.

ACCESS SARS-CoV-2 IgG QC SARS-CoV-2 IgG REF C58964 - Product Usage: use on the Access Family of Immunoassay Systems only.

DateDec 30, 2020
StatusTerminated
ReasonThe Quality Control (QC) card has a label error for the standard deviation (SD) values for one out of the two levels of QC.
Root causeLabeling mix-ups
ActionOn or about 01/04/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification to customer via mail or email informing them that the Recalling Firm has determined that the QC kit lot includes a Quality Control value card with an incorrect QC1 standard deviation (SD) value which may lead to a QC range that is too wide for the control level. Customers are instructed to: 1) Discard the incorrect QC value card included with the affected SARS-CoV-2 IgG quality control lot. 2) Use the Access SARS-CoV-2 IgG QC value card that is included with the Recall Notification letter in place of the disc
Quantity276 units
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, GA, IA, ID, IL, IN, KS,KY, LA, MD, MI, MN, MO, NC, ND, NY, OH, OK, OR, PA, Puerto Rico, SD, TX, WA, and WV. The countries of Andorra, Belgium, Brazil, Bulgaria, Canada*, Croatia, France, Germany, Hong Ko