Recall Z-1105-2021 ยท NGX Stimulator, Muscle, Powered, For Muscle Conditioning

BEMER INT. AG

Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), REF. 410100 (BEMER Classic-Set), REF. 420200 (B.BOX Professional), and REF. 410200 (BEMER Pro - Set)

DateJan 13, 2021
StatusTerminated
ReasonThere potential that electromagnetic fields generated in the therapy system may cause technical defects in a patient using an insulin pumps or a malfunction of a pacemaker.
Root causeLabeling Change Control
ActionOn January 13, 2021, Bemer USA sent a "Urgent Safety Information" letter to affected consignees via E-Mail. In addition to informing consignees about the recall the firm asked consignees to take the following actions: 1. Patients who wear an active medical implant that leads to stimulation (e.g., pacemakers, defibrillators, brain stimulators, muscle stimulators) should not use the BEMER therapy system unless a specialist physician carries out a personal risk assessment and subsequently expressly tells the patient that he/she can continue the treatment. If you are an affected patient, please
Quantity197,724 units
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151834BEMER Classic Set, BEMER Pro-Set
Bemer Int AG, Feb 22, 2017