Recall Z-0882-2021 ยท HRS Plate, Fixation, Bone

Biomet, Inc.

A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions of the clavicle including osteopenic bone.

DateJan 21, 2021 (initiated Dec 23, 2020)
StatusTerminated
ReasonThe instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill.
Root causeDevice Design
ActionOn 12/30/2020, Zimmer Biomet issued an Urgent Medical Device Recall to the consumer/user level, including any intermediate wholesale or retail consignees. All distributors will be notified via email. Hospital risk managers and surgeons, as well as distributors with product, will be notified via courier.
Quantity174 systems
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173767A. L. P. S. Clavicle Plating System
Biomet, Inc., Apr 6, 2018