Recall Z-0960-2021 ยท FAD Stent, Ureteral

Cook Inc.

Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr. 22-32cm -Used for retrograde injection during Extracorporeal Shock Wave Lithotripsy (E.S.W.L.) and leaving an indwelling ureteral stent Reference Part Number (RPN): 003700 GPN: G14837

DateJan 4, 2021
StatusTerminated
ReasonManufactured with incorrect material, which may result in the stent being stiffer and potentially susceptible to degradation could potentially lead to breakage/fragmentation.
Root causeNonconforming Material/Component
ActionCook issued Urgent Medical Device Recall Letter on Jan 4 , 2021 via email/mail. Letter states reason for recall, health risk and action to take: If affected products are currently indwelling in a patient, consider scheduling a follow-up with the patient to remove and/or replace the product, based on the status and preferences of each individual patient. 2.Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 3.Return the affected product(s) to Cook Medical w
Quantity42 units
DistributionFL, KY, MD, NS, NY, PA, TX Foreign: Canada, Switzerland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162109Kwart Retro-Inject Ureteral Stent
Cook Incorporated, Jun 7, 2017