Recall Z-0771-2021 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus

DateNov 24, 2020
StatusTerminated
ReasonSystem does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan
Root causeSoftware design
ActionCustomer notification letters will be sent starting December 14, 2020 to all affected customers via Customer Advisory Notice CT057/20/S. The technical solution, software versionVB20A_SP3, will be distributed to all affected customers free of charge via CT056/20/S. For questions or technical support, contact your local application specialists or your local service/sales organization at 1-800-888-7436.
Quantity3044
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103127SOMATOM DEFINITION AS
Siemens Medical Solutions USA, Inc., Mar 4, 2011
K121072SOMATOM DEFINITION FLASH
Siemens Medical Systems, Inc., May 8, 2012
K133589SOMATOM FORCE
Siemens Medical Solutions USA, Inc., Apr 17, 2014
K142955SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open,
Siemens Medical Solutions USA, Inc., Nov 24, 2015
K161196SOMATOM Drive
Siemens Medical Solutions USA, Inc., Aug 24, 2016
K173607SOMATOM Edge Plus
Siemens Medical Solutions USA, Inc., Mar 21, 2018