Recall Z-0969-2021 ยท FGE Stents, Drains And Dilators For The Biliary Ducts
Aomori Olympus Co., Ltd.
PR-V614M 2-LUMEN CANNULA, model no. PR-V614M - Product Usage: Used with endoscopes to inject contrast medium in the biliary or pancreatic duct.
| Date | Jan 4, 2021 |
| Status | Terminated |
| Reason | Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used. |
| Root cause | Under Investigation by firm |
| Action | On January 4, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that the identified list of Endo-therapy devices may have a defective seal which may compromise the sterility of the product. Customers were asked to take the following actions: 1. Immediately assess any product you have in stock to identify endotherapy products with affected lot numbers listed in the recall letter. Cease use of product and quarantine any affected product. 2. Call your Olympus customer service representative at 1-888-524-7266 to obtain a Returned Good |
| Quantity | 7,400,000 (global); 260,395 (US) |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K950729 | OLYMPUS PR-23Q DISPOSABLE CANNULA (BALL TIP) Olympus America, Inc., Mar 17, 1995 |