Recall Z-1133-2021 ยท HJO Biomicroscope, Slit-Lamp, Ac-Powered

Carl Zeiss Meditec AG

IOLMaster 700

DateDec 7, 2020
StatusTerminated
ReasonWhen using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen
Root causeSoftware design
ActionOn 12/11/2020, the firm sent a "FIELD SAFETY CORRECTION ACTION URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS informing them about a software error with software versions 1.90.2.09 and 1.90.8.06 that when used in conjunction with modality worklist functionality, is not performing as expected. These issues may lead to false display of patients in the patient manager screen, which can subsequently lead to the implantation of a wrong intraocular lens (IOL). The Recalling Firm is informing it customers that a software update needs to be installed on their device(s), or the modality wo
Quantity613 devices
DistributionU.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Austria, Australia, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, France, United Kingdom, G
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173771IOLMaster 700
Carl Zeiss Meditec, AG, Aug 24, 2018