Recall Z-1134-2021 ยท LKD Monitor, Carbon-Dioxide, Cutaneous

SenTec AG

SenTec Digital Monitor, Model SDM, is a portable stand-alone patient monitor indicated for continuous, noninvasive patient monitoring of carbon dioxide partial pressure (PCo2), oxygen partial pressure (POs), functional oxygen saturation (PsO2), and pulse rate (PR), using either a single digital sensor (V-Sign Sensor 2 or OxiVenT sensor).

DateDec 10, 2020
StatusTerminated
ReasonDevice requires the use of an Isolation Transformer in combination with the SDM in home care settings to comply with applicable safety standards.
Root causeDevice Design
ActionThe firm initiated their correction by letter to the direct accounts on 12/10/2020. The letter explains the problem and provides additional instruction manuals and a sticker to address the use of the product in the home care environment.
Quantity15 devices
DistributionIL, PA, FL, NJ, GA, MA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041548SENTEC DIGITAL MONITOR MODEL #SDM
Sentec, Inc., Aug 9, 2004
K101690SENTEC DIGITAL MONITOR, V-SIGN SENSOR AND V-SIGN SENSOR 2 WITH DIGITAL SENSOR ADAPER CABLE
Sentec AG, Dec 3, 2010
K151329SenTec Digital Monitor, OxiVenT Sensor, V-Sign Sensor, Staysite Adhesive Pad, V-STATS PC Software including V-CareNet
Sentec AG, Dec 17, 2015