Recall Z-0937-2021 ยท OWD Somatic Gene Mutation Detection System

Qiagen Sciences LLC

therascreen PIK3CA RGQ PCR Kit (23) US IVD product- aid clinicians in identifying breast cancer patients who may be eligible for treatment with PIQRAY (alpelisib) based on a PIK3CA Mutation Detected result. REF 873121

DateDec 11, 2020
StatusTerminated
ReasonKit may generate false Q546R mutation positive results caused by nonspecific molecular interactions within the Q546R higher than previously observed and described within the Instructions For Use.
Root causeProcess control
ActionQiagen US notified customers by Urgent Medical Device Correction letter via e-mail on 12/11/20. The e-mail additionally contains the letter as PDF. Letter states reason for recall, health risk and action to take: For patient samples where a PIK3CA Mutation Detected Result is obtained, disregard Q546R positive results. Cease to report patient samples where a Q546R mutation result is obtained as PIK3CA Mutation Detected. Information about individual target results is specifically obtained from the Individual target result column of the Rotor-Gene Assay Manager v2.1 result table. If multiple m
Quantity575 kits US
DistributionCA, IN, MI, MN, NC, NJ, NM, OH, TX
FDA recordOpen on fda.gov