Recall Z-0883-2021 ยท JQP Calculator/Data Processing Module, For Clinical Use
Roche Diagnostics Operations, Inc.
A calculator/data processing module for clinical use is an electronic device intended to store, retrieve, and process laboratory data Catalog Number: 07154003001
| Date | Dec 17, 2020 |
| Status | Terminated |
| Reason | A software error results in the unintentional removal of the serum-indices flag that would otherwise prevent the release of results. The Cobas Inifinity has an auto-verification feature to hold results for manual review when they meet specific criteria. The serum indices flag is used for serum indices tests that are performed to assess the quality of the sample (e.g. hemolysis, icterus, and lipemia). Normally, when the Cobas Infinity receives a result for a test that is serum indices-sensitive, it flags the result and the software then waits for the results of the serum indices tests before va |
| Root cause | Software design |
| Action | Roche issued Urgent Medical Device Correction, (UMDC TP-01176) via UPS Ground (receipt required) on 12/17/20. The UMDC TP-01176 will also post to the diagnostics.roche.com website. Letter states reason for recall, health risk and action to take: Complete the enclosed faxback form, TP-01177, and return it following the instructions on the form even if you are not using the Cross SI option. " If you are using the Cross SI option, refer to the Workarounds section of this Urgent Medical Device Correction (UMDC) until the Service Patch for your cobas infinity central lab is available for this issu |
| Quantity | 76 licenses |
| Distribution | US Nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.