ClearanceBrief
Recall Z-0892-2021 ยท MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Respironics California, LLC

Philips Respironics V60 Ventilator Part Number R1053618 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.

DateOct 30, 2020
StatusTerminated
ReasonVentilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
Root causeDevice Design
ActionOn 10/30/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the Ventilators may experience two potential issues with the battery (see table below for additional information). 1. The ventilator may exhibit a high priority Check Vent: Battery Failed alarm if the battery is allowed to over discharge. 2. The battery may fail to charge if the ventilator is stored for an extended period of time without being plugged into AC power. If the ventilator exhibits a high priority Battery Failed alarm while connected to AC power, the V60
Quantity3 systems
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, W
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082660V60 VENTILATOR, MODEL V8000
Respironics California, Inc., Jan 9, 2009