Recall Z-0933-2021 ยท LOD Bone Cement

Smith & Nephew, Inc.

VERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement which can be used for hip, knee, and shoulder implant fixation. VERSABOND AB includes the antibiotic- gentamycin.

DateDec 18, 2020
StatusTerminated
ReasonPotential heat and humidity exposure. The affected lot contained an atypical sandy, flaky or grainy consistency when mixing.
Root causeStorage
ActionA Recall notification letter titled, "Urgent Medical Device Recall Notice" was sent to consignees on 12/18/2020 via email and FedEx. The letter instructs the consignee to follow the instructions on the attached response form. The response form provides the following instructions: "1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions
Quantity5190 devices
DistributionWorldwide distribution - US Nationwide distribution in the state of OH and the countries of Dubai, Canada, Korea, Chile, Puerto Rico, Panama, and Costa Rica.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022688VERSABOND AB BONE CEMENT
Smith & Nephew, Inc., Apr 29, 2004