Recall Z-0656-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno

DateNov 30, 2020
StatusOpen, Classified
ReasonWhen the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
Root causeSoftware design
ActionSiemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers via AX067/20/P on November 30, 2020. All modifications and repairs are provided to affected customers free of charge.
Quantity5602
DistributionWorldwide-US Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018