ClearanceBrief
Recall Z-0858-2021 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

DateDec 10, 2020
StatusTerminated
ReasonGE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Root causeProcess control
ActionThe firm initiated their action by letter to the consignees on 12/10/2020. The letter explained the issue and stated that GE would inspect, and if required, correct the affected units. Questions and concerns may be directed to GE Healthcare Service at 1-800-437-1171 or the consignee's local Service Representative.
Quantity7 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121019OPTIMA NM/CT 640
GE Healthcare, Apr 19, 2012