Recall Z-0939-2021 ยท MVR Device, Anastomotic, Microvascular

Baxter Healthcare Corporation

FLOW COUPLER Monitor Flow Coupler devices are sold in 5 sizes (2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm). Implantable device for vascular use.

DateDec 28, 2020
StatusOpen, Classified
ReasonPotential for the battery to lose its ability to be recharged.
Root causeSoftware design
ActionWritten notification letter titled, "Urgent Medical Device Recall" was sent to all consignees on 12/28/2020 via first class mail. The letter instructs the consignee to locate all affected devices at their facility; contact Synovis MCA to arrange for product return for the firmware upgrade, power button update, and battery replacement, if applicable; complete the enclosed Baxter customer reply form and return it to Baxter by e-mailing it to fca@baxter.com, even if the consignees does not have any inventory; and if the consignee distributes this product to other facilities or departments within
Quantity109 units
DistributionDistributed nationwide to AZ, FL, OH, NY, GA, VA, MA, WI, MO, MT, WV, IL, NC, NJ, AR, CA, CO, CT, IA, ID, IN, KS, LA, MD, MI, MN, MS, NE, NH, NM, NV, OK, OR, PA, RI, WY, WA, VT, TX, TN, SC, SD
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191252Gem FlowCoupler System
Baxter/ Synovis Micro Companies Alliance, Inc., Jul 9, 2019