Recall Z-1075-2021 ยท DYB Introducer, Catheter
Argon Medical Devices, Inc
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19S, 1/19 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/5f straight catheter, Sterile EO,
| Date | Dec 10, 2020 |
| Status | Terminated |
| Reason | Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function. |
| Root cause | Process change control |
| Action | On 12/11/2020, Argon Medical Devices, Inc. distributed an "Urgent-Product Recall Notice" to all affected consignees via US mail. In addition to informing consignees about the recall device, the firm ask consignees to take the following actions: 1. Complete the response form and return it to us as quickly as possible. This will allow us to begin staging replacement product for you and minimize interruption to service. 2. All affected product should be returned to our Argon Athens facility using RGA#______, attention Arbee Cummings. The mailing address is as follows: RGA# ____, Argon Medic |
| Quantity | 247 units |
| Distribution | US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Ger |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K022634 | CO-AXIAL INTRODUCER NEEDLE Promex, Inc., Nov 6, 2002 |