Recall Z-0957-2021 ยท FMG Stopcock, I.V. Set
ICU Medical, Inc.
7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
| Date | Dec 2, 2020 |
| Status | Terminated |
| Reason | Due to the incorrect filter assembly. |
| Root cause | Process control |
| Action | On 12/02/2020, the firm's third party contractor send an "URGENT: MEDIAL DEVICE RECALL" Notification letter via traceable mailing to direct and indirect consignees informing them that specific lots of IV extension sets that may contain an incorrect filter assembly of a 0.2 micron filter instead of a 1.2 micron filter and the potential hazards/risks potentially cause delay in therapy or under delivery of medication. Customers are instructed to: 1) Discontinue the use and distribution of the affected product immediately. Check inventory and quarantine all affected product at their facility. 2) |
| Quantity | Total Kits = 7,500 for all four lots |
| Distribution | Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, CA, CO, CT, FL, ID, IL, LA, MA, MN, NC, NJ, NY, TN, TX, VA, WA and WV. The country of Belgium. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K964435 | PRIMARY IV SET Icu Medical, Inc., Apr 2, 1997 |