Recall Z-0955-2021 ยท FMG Stopcock, I.V. Set

ICU Medical, Inc.

60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).

DateDec 2, 2020
StatusTerminated
ReasonDue to the incorrect filter assembly.
Root causeProcess control
ActionOn 12/02/2020, the firm's third party contractor send an "URGENT: MEDIAL DEVICE RECALL" Notification letter via traceable mailing to direct and indirect consignees informing them that specific lots of IV extension sets that may contain an incorrect filter assembly of a 0.2 micron filter instead of a 1.2 micron filter and the potential hazards/risks potentially cause delay in therapy or under delivery of medication. Customers are instructed to: 1) Discontinue the use and distribution of the affected product immediately. Check inventory and quarantine all affected product at their facility. 2)
QuantityTotal Kits = 7,500 for all four lots
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AZ, CA, CO, CT, FL, ID, IL, LA, MA, MN, NC, NJ, NY, TN, TX, VA, WA and WV. The country of Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964435PRIMARY IV SET
Icu Medical, Inc., Apr 2, 1997