Recall Z-0801-2021 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Ortho-Clinical Diagnostics, Inc.

VITROS Automation Solutions, VAS, Catalog Numbers 6844300 & 6844301 - Product Usage: intended to automate pre-analytical and post-analytical sample processing in the clinical laboratory.

DateJan 7, 2021 (initiated Nov 20, 2020)
StatusTerminated
ReasonA software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.
Root causeSoftware design
ActionOn November 20, 2020, the firm sent Urgent Product Correction Notification letters to affected customers to inform them of the issue. Customers were reminded to follow product labelling regarding removing all tubes before performing maintenance or initializing the Aliquoter & Labeller Module. There should never be sample tubes present when initializing the Aliquoter & Labeller module. The firm's release of TCA Software version 4.7 is the planned mitigation for this issue. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Quantity40 units
DistributionWorldwide distribution - US Nationwide distribution including the states of CA, CO, MA, MD, NC, OH, TX, VA, VT, and Belgium, Bermuda, Canada, China, France, Italy, Japan, Mexico, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom.
FDA recordOpen on fda.gov